Neutronics Inc. is a world-class provider of gas analysis and gas handling technologies. Our three operating divisions, Neutronics Refrigerant Analysis, Ntron, and Biomarine Inc. provide state-of-the-art products and services for commercial, industrial, governmental, and military clients around the globe.

Ntron Oxygen Analyzers & Gas Monitors

Neutronics’ Ntron division is a worldwide supplier of high quality industrial Oxygen Analyzers & Gas Monitors.

We offer a complete range of engineered solutions, with capabilities that range from monitoring ultra-trace levels of oxygen in process streams at parts per billion, to monitoring and reporting toxic, flammable, and asphyxiate gases in ambient environments.

Let Ntron assess your gas measurement and control needs. We’ll provide reliable, and affordable control solutions.

Engineered Solutions
Complete, custom solutions designed to meet specific user application needs. We provide analysis, engineering, installation, and ongoing maintenance.

Ntron Model Oxytron 2000™ Process Oxygen Analyzer

model oxytron 2000The Ntron Oxytron 2000 is a full-featured industrial process control instrument for measuring oxygen in industrial process vessels and gas streams.

By continuously measuring oxygen, the Model 2000 provides increased safety margins in process manufacturing environments. It reduces inert gas consumption and VOC emissions by allowing manufacturers to use inert gas only as needed to maintain safe oxygen levels.

The Model 2000 offers low installation cost and comes ready to install with no additional hardware/software needed. Automated sensor calibration and system diagnostics enable low operating and maintenance costs.

The Oxytron 2000 provides the gas analysis component of the Neutronics Inerting Control System (ICS), the only proven system that automatically controls the level of oxygen in process manufacturing. Originally developed for one of the country’s largest chemical manufacturers, Neutronics ICS has been proven reliable in thousands of installations.

Specifications:

Sensor: Up to two galvanic-type, available in various models to meet specific application needs

Measurement Range: 0.01 to 25% Oxygen, fixed range

Display: 0.5″ 4-line LCD alphanumeric display, backlit; color-coded LED’s for system status, red=Alarm, yellow=Warning, green=Normal

User Interface: 3 menu-driven soft key operators

Response Time: T90<15 sec.

Accuracy: +/- 1% full-scale deflection

Signal Interface: (2) 4-20mA, Isolated, range configurable, Max load 550ohms

Alarms & Controls: User adjustable alarms with programmable relay mapping for Inerting Control, Calibration Control, Sample prefilter back flush, check loss-of-flow indicator, fault alarm, maintenance alarm, multiple O2 alarms

Additional inputs: (2) 4-20mA user configurable, (2) digital calibration trigger

Power: 110/220VAC, 50/60Hz, 30W

Enclosure Options: NEMA 4X Stainless Steel, NEMA 4/7 Explosion proof, standard 19″ rack-mount. Custom multi-module rack mount enclosures are also available.

Warranty Limited: One yr.

Features:

  • Dual channel operation, four analyzers in one – control two processes with one instrument
  • Real-time sensor trending and maintenance-prediction
  • Programmable automated maintenance features, including sensor calibration and sample flow check
  • Programmable analog and digital process control outputs
  • Two field instrument inputs with alarms and relays – temperature, pressure, PH, RH, and more
  • Available in a variety of enclosures – NEMA 4X wall mount, NEMA 7 Explosion Proof, panel mount, rack mount
Ntron Model Mini ICS™ Process Oxygen Analyzer

model mini icsThe Mini ICS™ is Ntron’s most affordable, off-the-shelf solution for measuring oxygen in industrial process vessels and gas streams.

By continuously measuring oxygen, the Mini ICS provides increased safety margins in process manufacturing environments. It helps reduce inert gas consumption and VOC emissions by allowing manufacturers to use inert gas only as needed to maintain safe oxygen levels.

The Mini ICS offers extremely low installation cost and comes ready to install with no additional hardware/software needed.

The Mini ICS can be used as a stand-alone instrument or as the analysis component in the Neutronics Inerting Control System (ICS), the only proven system that automatically controls the level of oxygen in process manufacturing. Originally developed for one of the country’s largest chemical manufacturers, Neutronics ICS has been proven reliable in thousands of installations.

Specifications:

Sensor: Galvanic, available in various models to meet specific application needs

Measurement Range: 0.1% to 50% Oxygen – auto or fixed range

Display: 0.75″ 7-segment LED alphanumeric display; color-coded LED’s for system status, Red=Low Flow / O2 Alarm, Yellow=Fault, Green=Normal

User Interface: 3 push-button operators; MODE, UP, DOWN

Response Time: T90<15 sec.

Accuracy: +/- 1% full-scale deflection

Signal Interface: 4-20 mA analog output, 0-1/5/10VDC analog output, RS-232 service port

Alarms & Controls: User-adjustable set points with relay outputs for inerting control solenoid valve, calibration solenoid valve, fault alarm, O2 alarm

Power: 110/220VAC 50-60Hz, or 12/24VDC (10W)

Enclosure Options: NEMA 4 Panel Mount, NEMA 4 and NEMA 4X Wall Mount, NEMA 4/7 Explosion Proof, 19″ Rack Mount. Custom multi-module rack mount enclosures are also available.

Warranty: Limited, one yr.

Ntron Model C-5 Handheld Percent Range Oxygen Analyzer with Internal Pump

model c-5The Ntron Model C-5 is a battery powered, handheld oxygen analyzer with internal pump that can quickly provide analysis of 0-25% oxygen with 0.1% resolution. Rugged and compact enough to be carried in a briefcase or toolbox, the versatile, inexpensive C-5 is equally suited on the production floor, in research labs, or for tuning combustion processes.

Fuel Cell Oxygen Sensor Technology
The C-5 oxygen sensor is a small diffusion limited, fuel cell that converts oxygen concentration by volume directly into low level electrical current. Using a capillary as the diffusion interface provides for a sturdier sensor with better temperature and pressure coefficients than membrane type oxygen cells. The size of a “C” cell flashlight battery, the C-5 oxygen sensor is inexpensive, disposable, and unaffected by most background gases.

Two Sampling Modes
The C-5 features two sampling modes. The timed ‘ON’ mode provides for repeatability in multiple samples and maximizes battery life. The C-5 draws a sample for 20 seconds and holds the display for 30 additional seconds. In the continuous ‘CAL’ mode, the C-5’s pump draws a sample and displays its oxygen concentration continuously.

Easy to Maintain
An integral moisture trap and line filter protect the oxygen sensor from potential contaminants. The inexpensive, disposable oxygen sensor calibrates with ambient air and is simple to replace. The C-5 is powered by 4 ‘AA’ batteries.

High Purity Welding
The C-5 is ideally suited for high-purity welding. An oxygen analyzer, like the C- 5, is the only way to accurately measure low concentrations of oxygen in purged tubing and pipes in high purity weldments.

Specifications:

Sensor: Electrochemical fuel Cell

Life: 9-16 mo.
Sensor Guarantee: 6 mo. from date of shipment
Calibration Gas: Ambient Air
Calibration Period: Weekly (typical)

Measurement Range: 25% oxygen by volume

Resolution: 0.1% oxygen

Display: 0.4” LED

Response Time: 90% of step change in less than 15 sec.

Intrinsic Error (accuracy): 0.2% oxygen

Repeatability: 0.1% oxygen

Warm-up: Instantaneous

Sampling: Stainless steel sample probe. Integral particulate and condensate filter

Enclosure:

Material: Molded High-Impact Plastic
Size: 5.5” x 3.5” x 1.5”
Weight: 1 lb. with sensor cell and batteries

Power: 4 AA alkaline dry cells; Low Battery Warning Light

Features:

  • Low Cost
  • Digital Display of 0 – 25% Oxygen
  • Calibrates in ambient air
  • Particulate and moisture filters
  • Battery-powered portable design
Ntron MGT4 Process Oxygen Transmitter/Display

mgt4The Model MGT4 is an economical 2-wire Transmitter/ Display for process oxygen measurement and control. It is available, off-the-shelf for measuring oxygen in industrial process vessels and gas streams.

The MGT4 can be combined with one of Neutronics’ standard process oxygen sampling systems for a complete process Oxygen measurement solution, or the user can provide their own Analyzer technology, using the MGT4 simply as a transmitter / display.

Specifications:

Sensor - Electrochemical

Weight - 1.8 lb. (0.8 kg)

Dimensions - 4.85” x 4.85” x 2.96” nominal (122 mm x 170 mm x 75 mm nominal)

Operating Voltage - 7 to 30VDC

Outputs - 4-20 mA scalable (over and under range diagnostic outputs)

Display -
 4-digit LCD

Operating Temp - 
-4º to 104º F (-20º to +40º C)

Features:

  • Compact Design
  • Certified Intrinsically Safe
  • Convenient local display for easy calibration and maintenance
  • Easy to use local 4-button operator interface
  • Standard analog 4-20mA output with fault indication
  • Can be factory-mounted on any Neutronics process oxygen sampling system
Ntron Model 1100 – Compact, Percent Range Oxygen Analyzer with Disposable Electrochemical Sensor

model 1100The Ntron Model 1100 is a compact, affordable modular analyzer for use in a variety of industrial applications. The unit’s measurement range extends from 0-1% up to 0-100% and its small footprint makes it easy to install.

The Model 1100’s electrochemical sensor is available with weak acid or alkaline electrolytes. The Ntron 1100’s low cost makes it attractive for OEMs and end-users.

Specifications:

Sensor: Disposable electrochemical cell, remote mount

Measurement Range: 0-1%/0-10%/0-25%/0-50%/0-100% auto range or fixed range

Display: 0.75” (19.05mm) LED digital display, Color coded LEDs for system status, Alarms 1&2= RED, system fault= YELLOW, OK and online= GREEN

Response Time: <15 seconds for 90% step change

Intrinsic Error (accuracy): At constant temperature: ±2% range over operating temperature: ±5%. Range when calibrated @ target concentration @ STP: ±1% range

Operating Temperature: 5-45º C

Signal Interface: RS-232 service port, 4-20mA negative ground, 0-10VDC range ID and choice of 0-1VDC, 0-5VDC or 0-10VDC analog oxygen signal

Alarm Outputs: Two alarms with adjustable form C relay outputs. Configurable for fail-safe or fail-alarm mode, ascending or descending trip. One system fault relay. Status LEDs are provided for each alarm.
Unit is provided with external cell which is mounted locally to the area being sampled.

Electrical Connections: Removable terminal blocks on back of analyzer

Power, Analyzer: 5W at 115/230 VAC50/60Hz, single phase or 12-30VDC (battery backup capability)

Environmental Ratings, Analyzer: Faceplate: NEMA type 4, IP66 Housing: NEMA type 1, IP20

Warranty: Electronics: One year, Sensor: CAG-21 six mo. / CAG 250E 2 yr.

Weight: Analyzer: Less than 2lb. (0.9kg)

Features:

  • Large, extra-bright LED display
  • Color coded status LEDs
  • Remote mount sensor
  • Electrochemical sensor available with weak acid or alkaline electrolytes
Ntron Model 2100 Parts Per Million (PPM) Oxygen Analyzer with
Disposable Electrochemical Sensor

model 2100The Ntron Model 2100 is a compact, versatile analyzer that features a disposable electrochemical cell, accessible from the front faceplate.

The small footprint and compact size makes the Ntron Model 2100 easy to install. Range is 0-10ppm/0-100ppm/ 0-1,000ppm/over range to 10,000ppm in auto range or fixed range modes.

The Model 2100 features two alarms with adjustable form C relay outputs. The user can configure the alarms for fail-safe or fail-alarm mode, ascending or descending trip order.

Specifications:

Sensor: Disposable electrochemical cell, front access

Measurement Range: 0-10PPM/0-100PPM/0-1,000PPM/ over range to 10,000PPM auto range or fixed range

Display: 0.75” (19.05mm) LED digital display, Color coded LEDs for system status, Alarms 1&2= RED, system fault= YELLOW, OK and online= GREEN

Response Time: 30 sec. for 90% response in ppm range

Intrinsic Error (accuracy): At constant temperature: ±2% range over operating temperature: ±5% range. When calibrated @ target concentration @ STP: ±1% range

Operating Temperature: 5-45º C

Signal Interface: RS-232 service port, 4-20mA negative ground, 0-10VDC range ID and choice of 0-1VDC, 0-5VDC or 0-10VDC analog oxygen signal

Alarm Outputs: Two alarms with adjustable form C relay outputs. Configurable for fail-safe or fail-alarm mode, ascending or descending trip. One system fault relay. Status LEDs are provided for each alarm.

Sample Port: 1/8″ NPT female ports

Electrical Connections: Removable terminal blocks on back of analyzer

Power, Analyzer: 5W at 115/230VAC 50/60Hz, single phase or 12-30VDC (battery backup capability)

Environmental Ratings, Analyzer: Faceplate: NEMA type 4, IP66 Housing: NEMA type 1, IP20

Warranty: Electronics and sensor: One yr.

Weight: Analyzer: Less than 2 lb. (0.9 kg)

Features:

  • Large, extra-bright, color-coded status LEDs
  • Front access sensor
  • Chemical resistant sensor
  • Removable terminal blocks
Ntron Model 3100 Parts Per Million (PPM) Oxygen Analyzer

model 3100The Ntron Model 3100 provides an affordable and versatile Oxygen Analyzer solution for a variety of applications including semiconductor manufacturing and solder/re-flow furnaces. The unit provides high-purity gas measurement in a range of 0.3ppm to 100% and super fast response time of air to ppm 02 in less than 10 seconds.

A small footprint and compact size makes the Ntron 3100 easy to install. The unit’s expected operational life is five to ten years.

The low-maintenance, long-life Zirconia sensor has virtually unlimited shelf life and will not dry out or freeze.

 

Specifications:

Sensor: Rapid Response Zirconia Sensor/Remote Mount

Measurement Range: 0-10PPM/0-100PPM/0-1,000PPM/0-10,000PPM/ 0-100% auto range or fixed range

Display: 0.75” (19.05mm) LED digital display, Color-coded LEDs for system status, Alarms 1&2= RED, system fault= YELLOW, OK and online= GREEN

Response Time: <10 sec. to PPM from air

Intrinsic Error (accuracy): At constant temperature: ±2% range

Over operating temperature: ±5% range When calibrated @ target concentration @ STP: ±1% range
Operating Temperature – 5 to 45º C

Signal Interface: RS-232 service port, 4-20mA negative ground, 0-10VDC range ID and choice of 0-1VDC, 0-5VDC or 0-10VDC analog oxygen signal

Alarm Outputs: Two alarms with adjustable form C relay outputs. Configurable for fail-safe or fail-alarm mode, ascending or descending trip. One system fault relay. One alarm/relay for sensor heater OK.

Sample Port: 1/8” compression type fittings on remote sensor module

Electrical Connections: Removable terminal blocks on back of analyzer

Power, Analyzer: 5W at 115/230VAC 50/60Hz, single phase

Environmental Ratings, Analyzer: Faceplate: NEMA type 4, IP66
Housing: NEMA type 1, IP20

Warranty: Electronics and sensor: One year

Weight: Analyzer: Less than 2 lb. (0.9 kg)

Features:

  • Compact size
  • Zirconia Sensor
  • Fast response time
Ntron Model 4100 Parts Per Billion (PPB)Oxygen Analyzer

model 4100The Ntron Model 4100 provides an affordable and versatile Oxygen Analyzer solution for a variety of applications. The unit provides high-purity gas measurement in a range of 100% to ppb 02 with a super fast response time.

A small footprint and compact size makes the Ntron 4100 easy to install. The unit’s expected operational life is five to ten years.

The low-maintenance, long-life Zirconia sensor has virtually unlimited shelf life and will not dry out or freeze.

 

Specifications:

Sensor: Rapid Response Zirconia Sensor / Remote Mount

Measurement Range: 0-10PPB/0-100PPB/0-1,000PPB/0-10,000PPB/ 0-100% auto range or fixed range

Display: 0.75” (19.05mm) LED digital display, Color coded LEDs for system status, Alarms 1&2= RED, system fault= YELLOW, OK and online= GREEN

Response Time: <10 min. to PPB from air

Intrinsic Error (accuracy): At constant temperature: ±2% range

Over operating temperature: ±5% range
When calibrated @ target concentration @ STP: ±1% range

Operating Temperature: 5 to 45º C

Signal Interface: RS-232 service port, 4-20mA negative ground, 0-10VDC range ID and choice of 0-1VDC or 0-5VDC or 0-10VDC analog oxygen signal

Alarm Outputs: Two alarms with adjustable form C relay outputs. Configurable for fail-safe or fail-alarm mode, ascending or descending trip. One system fault relay. One alarm/relay for sensor heater OK.

Sample Port: 1” VCR-type fittings on remote sensor module

Electrical Connections: Removable terminal blocks on back of analyzer

Power, Analyzer: 5W at 115/230VAC 50/60Hz, single phase or 12-30VDC (battery backup capability) Out

Environmental Ratings, Analyzer: Faceplate: NEMA type 4, IP66, Housing: NEMA type 1, IP20

Warranty: Electronics and sensor: One yr.

Weight: Analyzer: Less than 2 lb. (0.9 kg)

Features:

  • Compact size
  • Zirconia Sensor
  • Fast response for PPB measurements
Ntron Model 5100 0-100% Range Oxygen Analyzer

model 5100The Ntron Model 5100 provides an affordable and versatile Oxygen Analyzer solution ideal for portable and fixed gas generation standard process monitoring. The unit delivers high-purity gas measurement in a range from 0-1% through to 0-100% in an auto range or fixed range modes.

A small footprint and compact size makes the Ntron 5100 easy to install. The unit’s expected operational life is five to ten years.

The low-maintenance, long-life Zirconia sensor has virtually unlimited shelf life and will not dry out or freeze.

 

Specifications:

Sensor: Mini Zirconia Sensor

Measurement Range: 0-1%/0-10%/0-25%/0-50%/0-100% auto range or fixed range

Display: 0.75” (19.05mm) LED digital display, Color coded LEDs for system status, Alarms 1&2= RED, system fault= YELLOW, OK and online= GREEN

Response Time: <15 sec. for 90% step change

Intrinsic Error (accuracy):

At constant temperature: ±2% range
Over operating temperature: ±5% range
When calibrated @ target concentration @ STP: ±1% range

Operating Temperature: 5 to 45º C

Signal Interface: RS-232 service port, 4-20mA negative ground, 0-10VDC range ID and choice of 0-1VDC or 0-5VDC or 0-10VDC analog oxygen signal

Alarm Outputs:
 Two alarms with adjustable form C relay outputs. Configurable for fail-safe or fail-alarm mode, ascending or descending trip. One system fault relay.

Sample Port: 1/8” NPT female ports

Electrical Connections:
 Removable terminal blocks on back of analyzer

Power, Analyzer: 5W at 115/230VAC 50/60Hz, single phase or 10 Watts @ 12-30VDC (battery backup capability)

Environmental Ratings, Analyzer:

Faceplate: NEMA type 4, IP66
Housing: NEMA type 1, IP20

Warranty: Electronics and sensor: One yr.

Weight: Analyzer: Less than 2 lb. (0.9 kg)

Features:

  • Compact, easy-to-install modular analyzer with small footprint
  • Large, extra-bright, color-coded LED status display
  • Removable terminal blocks
  • Long-life Zirconia sensor
  • Can be provided with internal or external sensor
Ntron Model 7100 0 – 2,000 Parts Per Million (PPM) Range Oxygen Analyzer

model 7100The Ntron Model 7100 provides an affordable and versatile Oxygen Analyzer solution that’s ideal for high-purity gas manufacturing, high quality welding applications, and semiconductor manufacturing applications. The unit provides high-purity gas measurement in a range of 0 – 2,000 ppm.

A small footprint and compact size makes the Ntron 7100 easy to install. The unit’s expected operational life is five to ten years.

The low-maintenance, long-life Zirconia sensor has virtually unlimited shelf life and will not dry out or freeze.

 

Specifications:

Sensor: Mini Zirconia Sensor

Measurement Range: 0-10PPM/0-100PPM/0-1,000PPM/0-2,000PPM/auto range or fixed range

Display: 0.75” (19.05mm) LED digital display, Color coded LEDs for system status, Alarms 1&2= RED, system fault= YELLOW, OK and online= GREEN

Response Time: <15 sec. to 90% step change @ PPM level

Intrinsic Error (accuracy):

±3PPM @ 0-50PPM/±5PPM @ 51-500PPM/±5PPM @ 501-2000 PPM O2 @ STP when calibrated @ target concentration @ STP: ±1% range

Operating Temperature: 5-45º C

Signal Interface: RS-232 service port, 4-20mA negative ground, 0-10VDC range ID and choice of 0-1VDC or 0-5VDC or 0-10VDC analog oxygen signal

Alarm Outputs:
 Two alarms with adjustable form C relay outputs. Configurable for fail-safe or fail-alarm mode, ascending or descending trip. One system fault relay. One alarm/relay for sensor heater OK.

Sample Port:
 1/8” NPT female ports

Electrical Connections:
 Removable terminal blocks on back of analyzer

Power, Analyzer: 5W at 115/230VAC 50/60Hz, single phase or 12-30VDC (battery backup capability)

Environmental Ratings, Analyzer:

Faceplate: NEMA type 4, IP66
Housing: NEMA type 1, IP20

Warranty: Electronics and sensor: One yr.

Weight: Analyzer: Less than 2 lb. (0.9 kg)

Features:

  • Compact, easy-to-install modular analyzer with small footprint
  • Large, extra-bright, color-coded LED status display
  • Removable terminal blocks
  • Long-life Zirconia sensor
  • Can be provided with internal or external sensor
Ntron Model 7100P – Compact Portable Parts Per Million (PPM) Trace Oxygen Analyzer With Pump Drive

model 7100PThe Ntron Model 7100P is an affordable solution that’s ideal for spot-checking of high-purity applications. Multiple analog and RS-232 outputs provide flexibility and the unit is ready to use after only a 2-minute warm-up.

Model 7100P can be exposed to air and measure ppm values in seconds.

The 7100P is provided with a pump for sampling of gas streams. The flow meter indicates the rate of the sample through the pump and into the analyzer.

 

 

Specifications:

Sensor: Mini Zirconia Oxygen Sensor with Integral Heater

Measurement Range: 0-1,000PPM Oxygen (over-range to 1500PPM)

Sample Method: Pump Drive with sample flow meter (includes internal mounted bypass check valve)

Display: 0.75” LED digital display,

Resolution: XXXX.X Color coded LEDs for system status: Alarms 1&2=RED, System Fault=YELLOW, System OK & on-line=Green

Response Time: <15 sec. for 90% response due to step change in PPM Oxygen @ 1 SCFH flow

Accuracy: ±3PPM @ 0-50PPM Oxygen (±1ppm with 2nd point cal near measurement point) ±5PPM at 51 500PPM Oxygen ±10PPM at 501-1,000PPM Oxygen

Operating Temperature: 32º-113º F (0º-45º C)

Signal Interface:

Serial service port: Bi-directional RS-232 Computer Communications of the following: Oxygen Concentration, Alarm Settings/Status, Fault Status

Analog Outputs: Voltage 0-1VDC or 0-5VDC or 0-10VDC +/- 0.5% @ 1K input min.

Non-isolated: 4-20mA: +/- 0.5%, negative ground (loop is powered by analyzer) Maximum loop impedance 500Ω.

Sample Port: 1/8 inch compression tube fitting (1/8” tubing assembly included)

Warm-Up Time: Approximately 2 min. to reach thermal equilibrium with ambient temperature.

Power: 9-250VAC, 50/60 Hz, Single Phase 10W

Power Electrical Connections: 2 meter power cable, ON/OFF Switch

Storage Temperature: 15º to 50º C (5-122º F)

Humidity: 0-95% non-condensing

Sample Flow: 1.0 to 2.0 SCFH @ constant pressure

Warranty: Analyzer & sensor: 12 mo. from date of shipment under normal operation

Weight: Less than 5 lb.

Mechanical: Aluminum chassis with carrying handle on top and rubber feet

Features:

  • Long life sensor (3-5 yr) provides low maintenance costs
  • Universal power input (90-250VAC)
  • Large, super bright display LED display
  • Pump drive for easy extractive sampling
  • Local indication of gas flow
Ntron Pure N2 Parts Per Million (PPM) Oxygen Analyzer

pure n2The Ntron Pure N2 Oxygen Analyzer offers a cost effective solution to rapid and precise oxygen measurement. Ideally suited for spot checking of solder re-flow ovens, nitrogen sources and high purity welding, an oxygen analyzer, like the Pure N2, is the only way to accurately measure low concentrations of oxygen in purged tubing and pipes in high-purity weldments. The unit features rapid warm-up time and high sensitivity to low PPM levels of oxygen in a variety of background gases including Nitrogen, Argon, and Helium. Two alarms with independent relays may be set and controlled by the operator.

The Pure N2 provides fast response, high reliability, step change in 5 seconds, and rapid recovery from air to ppm levels. The Zr02 Sensor provides a range of 1PPM-100%. The Pure N2 is easy to own and maintain with no electrolytes to replenish or fuel cells to replace.

Compact and lightweight, the Pure N2 operates in process environments ranging from atmospheric pressure to vacuum. The keypad command system and LCD display assure ease of use and control of analyzer operation, data acquisition, and processing.

 

 

Specifications:

Sensor: Rapid response solid state Zirconia cell

Measurement Range: 0.1PPM to 100% 02. Analyzer auto-ranges between PPM and %. Indicator light on display specifies whether reading is PPM or %.

Response Time: 95% of step change in 5 seconds at 1 liter/min. sample flow rate (within range).

Intrinsic Error (accuracy): ±2% of RANGE

Operations:

Ambient Temperature: 32º to 113ºF (0º-45ºC)
Relative Humidity: 0-90% (non-condensing)
115/230 VAC ±20%
50/60 Hz(via power selector switch)

Warm-Up Time: Approximately 20 min.

Recovery Time:

100PPM <20 sec.
10PPM <5 min.
1PPM <15 min.

High Purity Pump: 316 SS head w/ Teflon™ Diaphragm.

Vacuum Capability: 12”Hg gauge,Flow: 2 LPM under normal process.

Dimensions: 13.3” H x 9” W x 14.75” L (338mm x 229mm x 375mm)

Features:

  • High-purity, heavy-duty pump with 316 SS pump heads and Teflon™ diaphragm
  • Microprocessor control with menu prompts
  • LED and LCD displays
  • Bi-directional RS-232 communications link
  • Analog outputs 0-1VDC or 0-10VDC
  • 4-20mA current output
  • Filter module kit
Ntron Model OA-1 Portable Parts Per Million (PPM) Oxygen Analyzer

model oa-1Ntron Model OA-1 is the industry benchmark for transportable oxygen analyzers for ultra-pure gas applications. Capable of analyzing PPM levels of oxygen in inert gases, the OA-1 delivers rapid response with high reliability to 1PPM O2.

The compact, lightweight design allows the OA-1 to meet spot check, QA/QC, and process control applications.

The user-friendly, keypad command system and LCD display assure ease of analyzer control. Microprocessor control and large LED display facilitate analyzer operation, data acquisition, and processing. The OA1 has two alarms with independent relays, which may be set and controlled by the operator.

Specifications:

Operating Conditions: Ambient Temperature: 32º to 113ºF (0º-45ºC) – Relative Humidity: 0-90% (non-condensing)

Range: 0.1PPM to 100% 02. Analyzer auto-ranges between PPM and %. Indicator light on display specifies whether reading is PPM or %.

Accuracy: ±2% of RANGE

Response Time: 95% of step change in 5 sec. at 1 liter/min. sample flow rate. (within range)

Recovery Time (from air to:): 100PPM <20 sec., 10PPM <5 min., 1PPM <15 min.

Operations: 10-60 PS1G. Flow Rate: .5 to 1.5 liters/minute, 115/230 VAC ±20%, 50/60 Hz. (via power selector switch)

Warm-Up Time: Approximately 20 min.

Dimensions: 8.12” H x 9” W x 13.55” L (206mm x 229mm x 244mm)

Weight: 10.5 lb (4.77 kg)

Features:

  • Bi-directional RS-232 communication link
  • Isolated or non-isolated 4-20MA current output
  • Analog outputs 0-1VDC or 0-10VDC
  • LED & LCD displays
Ntron Model OA-1S+ Parts Per Billion (PPB) Oxygen Analyzer

model oa-1sThe Ntron Model OA-1S+ is the industry benchmark for portable parts per billion Oxygen Analyzers for ultra-pure applications. The OA-1S+ is the only oxygen analyzer that can effectively measure from percent to parts-per-billion levels of oxygen, with very quick response.

Designed specifically for the demanding needs of the semiconductor industry, this portable analyzer also enables quick and effective checks of new piping welds, leaks in ultra-pure gas delivery systems, and verification of gas purity at point of use. The OA-1S+ can also serve as a cost effective back-up for other online process oxygen analyzers.

The unit features microprocessor controls with user-friendly menu prompts, and employs two Form C user-settable alarms.

Specifications:

Operating Conditions: Ambient Temperature: 32º to 113ºF (0º-45ºC), Relative Humidity: 0-90% (Non-Condensing)

Range: Auto-ranging. Indicator light specifies whether the reading is in PERCENT, PPM or PPB.

Response Time:

Change response:
95% of step change in 5 seconds at 1 liter/min. sample flow rate within PPM-% levels.
90% of step change in 30 sec. at PPB levels.

Recovery time: 
From room air to .1% <20 sec.
Low PPM levels <5 min.
Under 500 PPB <30 min.

Operations: Pressure= 10# to 60# PSIG. Flow rate 0.5 to 1.5LPM, 115/230VAC ±20%, 50/60Hz. (via power selector switch)

Warm-Up Time: Approximately 20 min.

Dimensions: 8.12” H x 9” W x 13.55” L (206mm x 229mm x 244mm)

Weight: 10.5 lb (4.77 kg)

Features:

  • Large LED digital display of oxygen concentration
  • Alphanumeric LCD displays mode of operation
  • Bi-directional RS-232 communication link
  • Analog outputs: 4-20MA and 0-1VDC
  • 316 S.S. electro polished tubing and fittings
  • Orbitally welded system
  • VCR-style connection

Gas Monitors

Ntron Model MGT3 – Intrinsically Safe Gas Detector

mgt3The Ntron Model MGT3 is a Certified Intrinsically Safe Gas Transmitter / Display for use in industrial gas monitoring applications. It can be configured to detect a variety of flammable and toxic gases as well as CO2 and Oxygen.

An LCD display and pushbutton operation make the Model MGT3 easy to operate and configure. The unit comes ready for direct integration with plant PLC or DCS. A plug-in sensor design makes maintenance fast and easy. The MGT is housed in a rugged, waterproof, and dustproof enclosure that features a tamper-proof cover.

Specifications:

Sensor: Pellistor, electrochemical, non-dispersive Infrared (NDIR)

Weight: 1.8 lb. (0.8 kg)

Dimensions: 4.85” x 4.85” x 2.96” nominal (122mm x 170mm x 75mm nominal)

Operating Voltage: 7 to 30 VDC

Outputs: 4-20 mA scalable (over and under range diagnostic outputs)

Display: 4-digit LCD

Operating Temp: -4º to 104º F (-20º to +40º C)

Features:

  • Intrinsically safe design
  • Easy installation and sensor change
  • Local display and pushbuttons for easy calibration
  • Standard 4-20mA output with fault indications
  • Flashing power-on indicator
  • Heavy-duty industrial NEMA 4X enclosure
Ntron Model MGT9 – All-in-One Explosion Proof Gas Detector & Controller

mgt9The Ntron Model MGT9 is a self-contained Gas Detector & Controller, for use in industrial gas monitoring and control applications. The unit is certified as “explosion proof”. It has three alarms, relays, and built-in local indicators and pushbuttons. It can be configured to detect over 150 flammable and toxic gases, as well as CO2 and Oxygen.

The Model MGT9 is capable of stand-alone control but comes ready for direct integration into plant PLC or DCS. It features a plug-in replaceable sensor for ease of maintenance.

Specifications:

Sensor: Pellistor, electrochemical, non-dispersive Infrared (NDIR)

Enclosure: NEMA 4/7 Explosion Proof

Display: 4-digit LCD

Outputs Analog:

4-20mA current loop
Digital: RS-232 serial
Three Form-C dry relay contacts

Operating Voltage: 12 to 24VDC

Operating Temp: -4º to 104º F (-20º to +40º C)

Dimensions: 4.80″ W x 5.59″ H x 2.95″ D
(122 mm x 142 mm x 75 mm)

Weight: 3.3 lb. (1.5 kg)

Features:

  • Heavy-duty Industrial NEMA 4/7 enclosure
  • Local display for easy calibration
  • Local magnetic pushbuttons for non-intrusive maintenance and calibration
  • Local LED Indicators
  • Standard 4-20mA output with three Alarm relays
Ntron Model SCA – All-in-One General Purpose Gas Controller

model scaThe Ntron Model SCA is a stand-alone Gas Detector & Controller for use in industrial and laboratory gas monitoring applications. Featuring a wide range of gas detection capabilities, it can operate as a stand-alone unit but comes ready for direct integration with plant PLC or DCS. It can be configured to detect over 150 flammable and toxic gases, as well as CO2 and Oxygen. An integral sensor is included but the SCA can operate through a remote mounted sensor as well.

The Model SCA features three built-in alarm strobes and an alarm horn. A large, backlit, LCD alphanumeric display makes for easy operation, and external push buttons enable fast calibration & configuration. The unit is housed in a rugged, waterproof, and dustproof steel enclosure, which includes a lockable door.

Specifications:

Sensor: Pellistor, Electrochemical, Non Dispersive Infrared (NDIR)

Weight: 15.72 lb (7.13 kg)

Dimensions: 9.8” W x 10.55” H x 5.25” D
(249 mm x 268 mm x 134 mm)

Display: Backlit alphanumeric LCD display

Operating Voltage: 120 or 240 VAC or 24 VDC
Optional battery backup (min. 8 hrs.)

Control Panel: Status LEDs Alarm (Red), Fault (Yellow), Power On (Green)

Outputs: 4 – 20mA signal or 0-5VDC
three Form C relays, 5A
Two 24VDC outputs for remote alarms
Internal audible and visible alarm

Features:

  • Heavy-duty Industrial NEMA 4X enclosure
  • One-piece stand-alone design for ease of installation
  • Red/Yellow/Green Strobe with Horn built-in by standard
  • Standard 4-20mA output with 3 Alarm relay
Ntron Model MCA – Multi-Channel Analyzer

model mcaThe Ntron Model MCA is a modular, multi-channel analyzer for use in industrial and laboratory gas monitoring applications where more than one gas detector is in use. Standard MCA systems can manage inputs from up to 32 remote Ntron MGT3 or MGT9 gas detectors, and can monitor a wide range of flammable and toxic gases, as well as CO2 and Oxygen. Custom systems are available. The MCA’s internal data logging has the capability of recording up to three months of data storage per channel. A large, backlit LED display and external push-button calibration and configuration make operation simple and straightforward.

The MCA features built-in audible and visible alarms, and comes housed in a rugged, waterproof and dustproof steel enclosure with a lockable door.

Specifications:

Input Type: 4-20mA (From MGT Transmitter)

Channels: Up to 8

Weight: 19.5 lb. (8 kg)

Dimensions: 18.7” W x 11.8” H x 10.0” D
(475mm x 300mm x 204mm)

Display: Backlit alphanumeric LED display

Operating Voltage: 120 or 240VAC or 24VDC
Optional battery backup (up to 17 hr.)

Control Panel:

Status LEDs- Alarm (Red), Fault (Yellow), Power On (Green)

Outputs- One 4-20 mA or (One) 0-5VDC per channel
Three Form C relays, 5A per channel
Four Form C relays, 8A (shared by all channels)
Three 24VDC outputs for remote alarms
Internal audible and visible alarms

Features:

  • Heavy-duty Industrial NEMA 4X enclosure
  • Large backlit LED ‘full graphics’ display
  • One-piece stand-alone design for ease of installation
  • Red/Yellow/Green Strobe with Horn built-in by standard
  • Standard 4-20mA output with 3 Alarm relays and common alarm output buss
Ntron Model PGA – Intrinsically Safe Portable Gas Monitor

model pgaThe Ntron Model PGA is a rugged, industrial 4-gas hand-held gas monitoring system. With up to four sensors per analyzer, it can be configured to detect flammable and toxic gases, as well as CO2 and Oxygen. Built-in audible and visible alarms include a 360-degree strobe and a 90dB horn.

The unit features an Intrinsically Safe design internal pump that contributes to the Model PGA’s Intrinsically Safe Certification.

Internal data logging can be downloaded through the PC interface facilitates to enable viewing or printing of data-logging info, instrument configuration, calibration (report printing), and programming. A convenient belt clip and hook is included, and rechargeable or dry battery packs are available.

Specifications:

Sensor: Pellistor, Electrochemical, Non Dispersive Infrared (NDIR)

Weight: 2.2 lb. (1 kg)

Dimensions: 4.25”W x 7.3” H x 1.97” D (108mm x 185mm x 50mm)

Display: 2×16 Character Backlit LCD

Operating Times: Rechargeable (12 hr.)
Dry Cell (24 hr.)

Alarm Outputs: Red flashing visual alarm, audible horn

Alarm Levels:

Flammables- 1 Level (high)
Oxygen- Two Levels (low, high)
Toxics- Three Levels (high, STEL, LTEL)

Features:

  • Heavy-duty industrial water-resistant enclosure with protective rubber boot
  • Intrinsically safe design with internal pump for hazardous area confined space entry
  • Data logging with PC utilities for easy management of gas reading records
  • ‘Remember High-reading’ mode for safe pre-entry into confined space

Ntron Engineered Solutions

Affordable, reliable, state-of-the-art Oxygen Analysis products are only half of what Ntron provides to our customers. The other half is our expertise in engineering complete monitoring and control solutions. Working with our customers, we create custom solutions designed to meet specific user application needs.

Whether it’s a simple, single-point monitoring requirement, or a complex, multi-point, multi-channel in-line process installation, Ntron can provide a comprehensive end-to-end solution.

Learn More:

Comprehensive Evaluation & Proposal Process

Every engineered solution Ntron creates, begins with a thorough analysis of our customer’s application parameters including evaluation of the measurement and control tasks required and the environment in which the monitoring will take place.

When evaluations are complete, our staff of engineers brings years of combined experience together to create an efficient, effective, and affordable proposal. We provide our customers with detailed solution proposals with comprehensive support documents and drawings for their evaluation. When all parties are in agreement that the solution will meet the need, and the equipment is made ready, the installation and commissioning can be scheduled.

Complete Product Solutions

Ntron can engineer and, manufacture, a complete system solution. Systems can include specialized subsystems for gas sampling and conditioning. Specialized adaptations can be made for monitoring in harsh/dynamic process environments. Systems can also be configured to include self-diagnostic and self-maintenance capabilities as well as user configurable alarm/intervention features.

Start-up & Training Services

As part of our engineered solutions, we offer complete start-up and commissioning services. Neutronics field service engineers provide guidance on the proper installation and start-up preparations to help ensure a smooth and successful start-up.

To ensure that the system operates as designed, we provide on-site training in system operation, maintenance, troubleshooting, and safety. Providing hands-on review of the equipment along with in-depth manuals for employees.

Ongoing Support Systems

A successful start-up isn’t the end of Ntron’s commitment to the customer. We can provide ongoing field evaluation and application engineering services as well as preventative maintenance, 24-hour emergency service, and training seminars for personnel.

Other on-site services include equipment inspection, cleaning, and calibration by Neutronics field service engineers. Based on the customer’s specific applications, we can recommend custom periodic service and maintenance schedules.

Spare Parts & Supplies

Ntron provides quick access to expendables and replacement/service parts and supplies. We maintain in-stock inventory for most application requirements.

 

Upgrades

When the need arises, Ntron can engineer and implement system upgrades, incorporating the latest hardware and software improvements to enhance or expand the capabilities of existing equipment. With the same user-intimate approach that we foster for any of our engineered solutions, we will work with your existing equipment, and design an upgrade program that fits your unique needs.